Description
Gastec detector tube No. 6 measures water vapour (H2O) in air from 0.5 to 32 mg/L. Break both tips off the tube, fit it to a Gastec GV-100 or GV-110 hand pump and draw the number of pump strokes listed for the expected range. The reagent changes from yellowish green to purple, and the stain shows the gas concentration. Read the concentration directly on the scale printed on the tube. 10 tubes per box give 10 tests.
Water vapour readings help check moisture in compressed air, process gas and gas dehydration lines. A tube gives a spot reading and does not replace a certified gas monitor where your site procedure requires one. Use only Gastec tubes with a Gastec pump, break the tips away from your eyes, and never handle broken glass or reagent with bare hands.
Features
- Measures water vapour (H2O) from 0.5 to 32 mg/L
- Colour change yellowish green to purple
- Direct-reading scale printed on the tube
- For use with a Gastec GV-100 or GV-110 hand pump
- 10 tubes per box, 36-month shelf life
- Each production lot is tested and calibrated by the manufacturer
Specifications
| Brand | Gastec |
|---|---|
| Manufacturer part number (MPN) | 6 |
| Big Papa SKU | BP-L-IG6 |
| Type | Colourimetric detector tube |
| Gas measured | water vapour (H2O) |
| Measuring range | 0.5–32 mg/L |
| Pump strokes and sampling time | 0.5–1 mg/L: 2 strokes (200 mL) in 1.5 min; 1–18 mg/L: 1 stroke (100 mL) in 45 s; 18–32 mg/L: ½ stroke (50 mL) in 30 s |
| Colour change | yellowish green to purple |
| Detection limit | 0.1 mg/L (2 strokes) |
| Operating temperature | 10 to 40 °C (50 to 104 °F) |
| Temperature correction | Required (see the instruction manual) |
| Tubes per box | 10 |
| Shelf life | 36 months |
| Model No. | 6 |