Gastec NH3 Detector Tube No. 3L, 0.5–78 ppm, Box of 10


Price:
Sale priceCFA70,000 XOF

Shipping & returns

  • Shipping: orders are processed Monday to Saturday. Delivery times and fees depend on the destination; international destinations are quoted on request.
  • Returns: eligible items can be returned within 30 days of receipt, unused and in their original packaging, with proof of purchase. Contact us first: returns sent without a prior request are not accepted.
  • Not returnable: perishable goods, custom or special orders, personal-care products (such as beauty products), hazardous products, flammable liquids or gases, promotional items and gift cards.
  • Please inspect your order on receipt and contact us right away if an item is damaged, defective or incorrect.

Description

Gastec detector tube No. 3L measures ammonia (NH3) in air from 0.5 to 78 ppm. Break both tips off the tube, fit it to a Gastec GV-100 or GV-110 hand pump and draw the number of pump strokes listed for the expected range. The reagent changes from pink to yellow, and the stain shows the gas concentration. Read the concentration directly on the scale printed on the tube. 10 tubes per box give 10 tests.

Ammonia is used as a refrigerant and in fertilizer and chemical processes, and leaks are a hazard in plants, cold stores and vessels. Use only Gastec tubes with a Gastec pump, break the tips away from your eyes, and never handle broken glass or reagent with bare hands.

Features

  • Measures ammonia (NH3) from 0.5 to 78 ppm
  • Colour change pink to yellow
  • Direct-reading scale printed on the tube
  • For use with a Gastec GV-100 or GV-110 hand pump
  • 10 tubes per box, 36-month shelf life
  • Each production lot is tested and calibrated by the manufacturer

Specifications

BrandGastec
Manufacturer part number (MPN)3L
Big Papa SKUBP-L-IG3L
TypeColourimetric detector tube
Gas measuredammonia (NH3)
Measuring range0.5–78 ppm
Pump strokes and sampling time0.5–1 ppm: 2 strokes (200 mL) in 1.5 min; 1–30 ppm: 1 stroke (100 mL) in 45 s; 30–78 ppm: ½ stroke (50 mL) in 30 s
Colour changepink to yellow
Detection limit0.2 ppm (2 strokes)
Operating temperature0 to 40 °C (32 to 104 °F)
Relative humidity0 to 90 %
Temperature correctionRequired (see the instruction manual)
Tubes per box10
Shelf life36 months
Model No.3L